AXIOM Artis System dBA using software version VB31D with 30x40 cm2 flat detectors. AXIOM Artis is a angiography system developed for single and biplane diagnostic imaging and interventional procedures.

FDA Recall Z-0313-2009 · Posted 2008-12-09

Recall StatusTerminated
Product CodeIZI
Date Initiated2008-08-25
Date Terminated2010-01-26
Product Quantity6 units

Reason for Recall

System may switch to emergency fluoroscopy mode.

Firm Action

A Customer Safety Advisory Letter dated 8/20/08 was issued with Update Instructions AX045/08/S. The letter informed the customer of the potential issue and provided instruction to avoid its occurrence. A software update (VB31E) was released to correct the issue.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution including states of AR, CA, CO, FL, IA, IL, IN, KS, KY, MI, NY, TX, VA, and WY.