SimpliRED D-Dimer, Product Code DSRK4 The SimpliRED D-dimer assay is a rapid, qualitative test for the detection of cross-linked fibrin degradation products containing the cross-linked D-diner site in human whole blood
FDA Recall Z-0314-2017 · Posted None
Advertisement
Recalling FirmBBI SOLUTIONS OEM LTD
Recall StatusTerminated
Product CodeGHH
Date Initiated2016-06-01
Date Terminated2018-06-14
Product Quantity420 units (183 units quarantined)
Reason for Recall
A component of the kit is showing a sensitivity performance detected in stability testing at 15 months of an 18-month shelf life. Qualitative results may be affected. False negative results are possible which may incorrectly influence the clinical diagnostic decision.
Firm Action
BBI Solutions sent an Urgent Medical Device Recall letter dated June 30, 2016, to all affected customers. Customers were advised to immediately discontinue use of the product, consider the requirement for review of reported test results, complete and return the customer acknowledgement form to BBI Solutions and to complete and return the customer reconciliation form to the distributor. Customers who have forwarded product to other sites, should inform those customers of the product recall and provide them with a copy of the notice. Customers with questions were instructed to call Tel: +44(0)1382 564093.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US distribution to CT and IN