Terminated

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required); Model 801764. (Not distributed within the United States).

FDA Recall Z-0315-2007 · Posted 2006-12-27

Recall StatusTerminated
Product CodeDTQ
Date Initiated2004-05-01
Date Terminated2009-07-14
Product Quantity5

Reason for Recall

The gas system may fail calibration prior to use due to incorrect gas system flowmeter software.

Firm Action

U.S. consignees were visited and corrections made between May 2004 and November 2005. Foreign affiliates were instructed to make corrections on international units, but the firm has not received confirmation yet that these units have been corrected.

Root Cause

Process control

Distribution Pattern

Worldwide distribution, including USA, Belgium, Germany, Japan, Korea and Republic of China.