Terminated

ProScreen 10 Drug Panel Cassette Test, Item No. CPSP-10PPX; ProScreen 10 Panel Cup, Item No. PSCup-10MMO-300, PSCup-10M, PSCupA-10MO, PSCUp-10AB; ProScreen 10 Panel Cup with Adulterants, Item No. PSCupA-10AM; ProScreen 10 Panel Dip Card, Item No. PSD-10MMO-300, PSD-10MOX, PSD-10M, PSD-10CMO3, PSD-10APO-300; ProScreen 10 Panel Dip Card w/Adult, Item No. PSDA-10MO, PSDA-10MMO-300; ProScreen 11 Panel Dip Card, Item No. PSD-11MT; ProScreen 12 Drug Cup w/Adulteration, Item No. PSCupA-12TBU; ProScreen 12 Panel Cup w/Adult., Item No. PSCupA-12M, PSCupA-12PP; ProScreen 12 Panel Dip Card, Item No. PSD-12BUP ProScreen 5 Panel Dip Card, Item No. PSD-5MB; ProScreen 5 Panel Dip Card w/Adult, Item No. PSDA-5MB-300; ProScreen 6 Panel Cassette w/Adult (AU), Item No. PSPA-6MBAU; ProScreen 6 Panel Cup w/Adult (AU), Item No. PSCupA-6MBAU; ProScreen 6 Panel Dip Card, Item No. PSD-6MBO, PSD-6MTDBO-300, PSD-6MB, PSD-6BUPO; ProScreen 7 Panel Dip Card, Item No. PSD-7MO, PSD-7M; ProScreen 8 Panel Dip Card, Item No. PSD-8P; ProScreen 8 Panel Dip Card w/Adult, Item No. PSDA-8P; ProScreen 9 Panel Dip Card, Item No. PSD-9P; ProScreen CLIA Waived Cup 6 Drugs w/Adul, Item No. PSCupA-6MB-W; ProScreen CLIA Waived Cup with 6 Drugs, Item No. PSCup-6BO-W; ProScreen CLIA Waived Cup with 8 Drug, Item No. PSCup-8P-W; ProScreen Cup, Item No. PSCup-6MB-W, PSCupA-6MB-W, PSCup-6BO-W, PSCup-8P-W, PSCUp-10AB, PSCup-10MMO-300; ProScreen Cup CLIA with 6 Panel Drug, Item No. PSCup-6MB-W; These Drugs of Abuse Tests are one-step immunoassays for the qualitative detection of multiple drugs and drug metabolites in human urine. The configurations of these Drugs of Abuse Tests consist of a combination of drug strips, including Benzodiazepine (BZO). They are used to obtain a visual, qualitative result and are intended for professional and point of care use.

FDA Recall Z-0315-2016 · Posted None

Recalling FirmAmeditech Inc
Recall StatusTerminated
Product CodeJXM
Date Initiated2015-10-20
Date Terminated2016-11-09
Product Quantity31,530 units

Reason for Recall

Ameditech is recalling drug abuse tests because they have shown reduced reactivity.

Firm Action

Ameditech Inc. sent an Urgent Medical Device Recall Letter to al affected customers on October 20, 2015, to inform them that Ameditech is recalling affected lots of Ameditech's Drugs of Abuse Tests because they have shown reduced reactivity for up to three of the nineteen Benzodiazepine (BZO) compounds. The letter informs customers of the actions to be taken. Customers were instructed to complete and return the customer verification form to (858) 677-0243 or scan and email to [email protected]. For questions regarding this recall call 858-535-1968.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US (nationwide) and Internationally to Australia.