Posterior Acetabular Retractor-Left with Light Cord Bracket, Model #BIO-1-1050-A, and Posterior Acetabular Retractor-Right with Light Cord Bracket, Model #BIO-1-1051-A, packaged 1 retractor (left or right retractor) per plastic bag. The firm name on the label is Whiteside Biomechanics, Inc., St. Louis, MO.

FDA Recall Z-0316-2007 · Posted 2006-12-27

Recall StatusTerminated
Product CodeGAD
Date Initiated2006-09-15
Date Terminated2008-06-11
Product Quantity10 retractors (5 right, 5 left)

Reason for Recall

The soft tissue pin in the retractor can break off and become embedded in the tissue.

Firm Action

The recalling firm was notified of the problem via a telephone call from the surgeon on 9/15/06 and the remaining retractors were placed in quarantine by the hospital. The recalling firm requested return of the retractors. The recalling firm immediately visited the one remaining hospital on 9/15/06 and picked up their inventory of six retractors.

Root Cause

Other

Distribution Pattern

Distribution was made to two hospitals, one in MD and the other in MO.