28CM DECATHLON GOLD FUNNEL TIP STD KIT Models: DFC23SH28 and DFC23SH28-K DECATHLON GOLD FUNNEL TIP 15.5 Fr polyurethane twin lumen Heparin coated hemodialysis catheter with separated funnel tips, cuff and side holes for chronic hemodialysis and apheresis. Kit contains catheter with stylet, 18 gauge introducer needle, 12cc syringe, J/straight 0.038 inch guidewire, 12Fr dilator, 14Fr dilator, 16Fr peel-away sheath/dilator, #11 scalpel, tunneling stylet, (4) gauze sponges, wound dressing, (2) injection sealing caps, and warning label. Latex free.

FDA Recall Z-0316-2010 · Posted 2009-11-17

Recall StatusTerminated
Product CodeMSD
Date Initiated2009-10-23
Date Terminated2010-06-09
Product Quantity1424 units

Reason for Recall

There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira

Firm Action

Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - Nationwide, Government. Foreign: Canada, France.