Safil QUI + UNDY 5/0 (1) 45CM DSMP13 Suture
FDA Recall Z-0316-2017 · Posted None
Advertisement
Recalling FirmAesculap Implant Systems LLC
Recall StatusTerminated
Product CodeGAM
Date Initiated2016-08-01
Date Terminated2018-02-16
Product Quantity180 units
Reason for Recall
There is a potential for the needle to detach from the suture thread during surgery. If the needle becomes detached the defective suture will need to be replaced causing the wound to be sutured a second time.
Firm Action
An Urgent Medical Device Recall Notification Letter was mailed to customers on August 11, 2016 asking them to remove and return the affected product immediately.
Root Cause
Process control
Distribution Pattern
Distributed in Arizona
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.