Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Ceres Self-Drilling Variable Screw 4.0 x 12mm The Ceres screws are self-drilling variable screws that are intended to be used with the Ceres-C Stand-Alone Cervical System.

FDA Recall Z-0316-2018 · Posted None

Recalling FirmAmendia, Inc
Recall StatusTerminated
Product CodeOVE
Date Initiated2017-07-07
Date Terminated2019-10-15
Product Quantity40 screws

Reason for Recall

Incorrect direct marking and color identification for the affected lot of Ceres Self-Drilling Variable Screws. Although the laser mark and color identify the screws as 12 mm in length, some screws in the lot measure 14 mm in length.

Firm Action

The firm, Amendia, sent an "URGENT: MEDICAL DEVICE RECALL" letter to its Customers. The letter described the product, problem and actions to be taken. The customers were Instructed to examine inventory and quarantine affected products, notify customers if affected products were further distributed, coordinate the return of affected products back to Amendia by emailing: [email protected] or call 770-575-5236 or jwilliamson@amendia or call 770-575-5369, and to complete and return the Response Form. If there are questions, Amendia Customer Service can be contacted at 770-575-5224.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Distributed to states of: AL, FL, GA, NY, TX, and VA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑