Fitmore Hip Stem, hip prosthesis, various sizes, Item Nos. 0100551102 0100551103 0100551104 0100551105 0100551106 0100551107 0100551108 0100551109 0100551110 0100551111 0100551112 0100551113 0100551114 0100551201 0100551202 0100551203 0100551204 0100551205 0100551206 0100551207 0100551208 0100551209 0100551210 0100551211 0100551212 0100551213 0100551214 0100551301 0100551302 0100551303 0100551304 0100551305 0100551306 0100551307 0100551308 0100551309 0100551310 0100551311 0100551312 0100551313 0100551314 0100551401 0100551402 0100551403 0100551404 0100551405 0100551406 0100551407 0100551408 0100551409 0100551410 0100551411 0100551412 0100551413 0100551414

FDA Recall Z-0320-2020 · Posted None

Recalling FirmZimmer Biomet, Inc.
Recall StatusTerminated
Product CodeLZO
Date Initiated2019-10-10
Date Terminated2020-05-19
Product Quantity5,459,583 total devices

Reason for Recall

Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).

Firm Action

The firm distributed Urgent Medical Device Correction letters to customers on October 10, 2019. The letter informed customers that the previous version of low density polyethylene (LDPE) bag could potentially adhere to the surface of metal devices. Please note, this is separate from the previous recall for highly polished implants packaged in the previous version of the LDPE bags. Customers were asked to take the following actions: 1. Review this notification and ensure that pertinent personnel are aware of the contents. 2. Ensure that backup devices for the affected products are available, as per standard orthopedic procedure planning, to reduce the likelihood of an extension of surgery time. 3. Examine devices thoroughly for adherent LDPE bag residue prior to use. Use a backup device if LDPE bag residue is detected. 4. Return any devices which display adherent LDPE bag residue to Zimmer Biomet, consistent with the standard instructions for use regarding the disposition of damaged components. 5. Complete the Certificate of Acknowledgement and return to the firm. 6. Retain a copy of the acknowledgement form with your field action records in the event of a compliance audit of your facility's documentation. If you have further questions or concerns regarding this recall, please call the firm's customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected].

Root Cause

Packaging change control

Distribution Pattern

Distributed nationwide.