Terminated

Streamline MIS Navigation System Taps, Catalog numbers 04-CANNDRILL-MDN 04-CANNTAP-40-MDN 04-CANNTAP-45-MDN 04-CANNTAP-55-MDN 04-CANNTAP-65-MDN 04-CANNTAP-75-MDN 04-CANNTAP-85-MDN 04-CTAP-TT-65-MDN 04-S-FINDER-MDN 05-SCREWINS-CL-MDN 05-SIMPDRIVER-MDN

FDA Recall Z-0320-2021 · Posted None

Recall StatusTerminated
Product CodeOLO
Date Initiated2020-09-11
Date Terminated2024-05-23
Product Quantity143 units

Reason for Recall

The Streamline MIS Navigation System taps are not compatible with the Streamline MIS tap sleeve.

Firm Action

On September 11, 2020, Pioneer Surgical sent out an advisory notice to their distributor (Surgalign) as a voluntary correction with no removal of devices from the field. However, after the advisory notice was distributed, a different issue related to the Streamline MIS Navigation taps becoming bound within Medtronic's Stealth Navigation System NavLock Tracker was reported. As a result, a second notification was sent out on September 24, 2020, changing the action to a voluntary recall. All instruments from the Streamline MIS Navigation System (as well as the Streamline TL Navigation System) were requested to be removed from the field. Please see RES 86559 for details on the September 24, 2020, recall. Customers are asked to do the following: -Remove all affected product from user sites and inventory locations -Return affected product to RTI Surgical, using RGA number - 63000897, along with a completed copy of the Acknowledgement and Receipt Form -Immediately notify any of your customers who have received the affected devices by providing them with a copy of the recall letter and response form for completion when returning affected devices.

Root Cause

Device Design

Distribution Pattern

Product distributed to Surgalign, a distributor located in Michigan.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →