Hillrom Welch Allyn HScribe Holter System; ambulatory monitoring
FDA Recall Z-0320-2025 · Posted 2024-11-07
Advertisement
Recalling FirmBaxter Healthcare Corporation
Recall StatusOpen, Classified
Product CodeMLO
Date Initiated2024-10-11
Product Quantity599 units
Reason for Recall
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Firm Action
Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/11/2024 via USPS First Class Mail. The notice explained the issue, potential hazard to the patient, and provided instructions for the to the system IT administrator to correct the issue which is already available in the current version of software.
Root Cause
Software design
Distribution Pattern
Worldwide distribution.