Terminated

HeartStart Pads Pediatric Plus, Pediatric Plus Multifunction Defibrillation Electrode Pads Model M3717A

FDA Recall Z-0321-06 · Posted 2005-12-23

Recall StatusTerminated
Product CodeMLN
Date Initiated2005-11-18
Date Terminated2006-07-12
Product Quantity253 CASES (5/CS)

Reason for Recall

The Apex/Anterior and the Sternum/Posterior pad labels are located in reverse position to the wire and connector

Firm Action

Philips Medical sent the recall notification letter 11/30/05 via FedEx

Root Cause

Other

Distribution Pattern

Nationwide Foreign: Australia, Canada, France, germany, Spain,, Sweden, Switzerland, UK