Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
FDA Recall Z-0321-2016 · Posted None
Advertisement
Recalling FirmMedtronic Sofamor Danek Usa, Inc - Dallas Distribution
Recall StatusTerminated
Product CodeHBB
Date Initiated2015-10-21
Date Terminated2016-06-24
Product Quantity9 pieces
Reason for Recall
The wire pass hole is not present on the tools.
Firm Action
A recall letter was mailed to consignees on October 21, 2015.
Root Cause
Process control
Distribution Pattern
Distributed in the states of California, New York, and Texas and the country of Belgium.