Terumo Advanced Perfusion System, 1 Roller Pump, 6 inch diameter; Catalog No. 801041, Terumo Cardiovascular Systems, Ann Arbor, MI 48103

FDA Recall Z-0322-2008 · Posted 2007-12-01

Recall StatusTerminated
Product CodeDTQ
Date Initiated2007-09-10
Date Terminated2008-02-23
Product Quantity76

Reason for Recall

Failure to Operate: The pump may fail to operate, and be difficult to hand crank, due to an internal electrical short condition.

Firm Action

Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, which provided them guidance in the event the problem occurs and advised them they would be contacted to schedule replacement of the circuit boards.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution-USA including states of Alabama, Arizona, Florida, Illinois, Indiana, Mississippi, Nebraska, North Carolina, Pennsylvania, and Virginia, and countries of Belgium, Hong Kong and United Arab Emirates.