Terumo Advanced Perfusion System 1 Integrated Centrifugal System Control Unit; Catalog Number 801046, Terumo Cardiovascular Systems, Ann Arbor, MI 48103

FDA Recall Z-0323-2008 · Posted 2007-12-01

Recall StatusTerminated
Product CodeDTQ
Date Initiated2007-09-10
Date Terminated2009-07-07
Product Quantity120

Reason for Recall

Failure to operate: The unit display may go blank and reset when the start button is pressed, due to an internal electrical short, resulting in a centrifugal pump that fails to start as expected.

Firm Action

Consignees were notified via an Urgent Medical Device Correction letter dated 9/10/07, given instructions to follow if the problem were to manifest itself, and informed the firm would contact them to schedule an upgrade.

Root Cause

Component change control

Distribution Pattern

Nationwide.