Reprocessed Pressure Tourniquet Cuff (PTC) Tourniquet cuffs are single- or dual-bladder inflatable cuffs connected to a tourniquet system via a hose assembly. When wrapped around a limb and inflated, tourniquet cuffs apply an adequate amount of pressure on the arterial blood flow in a limb to create a bloodless surgical field. Tourniquet cuffs are available in a variety of sizes to accommodate a wide range of limb circumferences.
FDA Recall Z-0324-2017 · Posted None
Advertisement
Recalling FirmStryker Sustainability Solutions
Recall StatusTerminated
Product CodeKCY
Date Initiated2016-09-19
Date Terminated2017-06-01
Product Quantity1,695,760
Reason for Recall
Issue with the process for detection of leaking Pressure Tourniquet Cuffs (PTC). Leaking PTC devices may result in excessive blood loss requiring surgical or medical intervention or systemic leakage of local anesthetic during Bier Block requiring medical intervention.
Firm Action
Stryker Sustainability Solutions sent an Customer Notification letter September 19, 2016.
Customers were instructed to discontinue use of the affected products. The letter stated to follow the link below to complete the Recall Effectiveness Check Form and indicate if any devices containing expiration dates from May 2016-May 2018 remain in inventory. This form must be completed even if no affected product is found.
Use the following reference number as you complete the form: [Ship to Account #]
If the form indicates that any affected devices remain in inventory, a prepaid shipping label will be issued for the return of the product. Credit for all affected devices returned will be issued to the facility.
.Adverse reactions or quality problems experienced with the use of this product may be reported to:
o Stryker Sustainability Solutions Complaint Hotline: +1(888) 888-3433 X5555
o http://www.stryker.com/productexperience/
o The FDAs MedWatch Adverse Event Reporting program either online, by
regular mail, or by fax.
Root Cause
Device Design
Distribution Pattern
U.S. distribution nationwide. No foreign distribution