Recall StatusTerminated
Date Initiated2017-11-15
Date Terminated2019-11-04
Product Quantity511
Reason for Recall
Potential for cracked luer hubs
Firm Action
On November 15, 2017 the firm sent letter to their customers indicating the following:
The purpose of this communication is to inform you of a potential leakage scenario with the S.M.A.R.T. Flex Biliary Stent System. The leak may occur due to potentially cracked hubs in some units from within the 29 lots listed in this letter.
Inspection details:
As stated in the Instructions for use, users are warned that the device should not be used if the outer sheath cannot be flushed. Inspect for leakage during preparation of the device. Leakage may be more likely in the affected product due to cracks in the luer. At time of actual flushing, if there is no leakage observed, the product may be safely used.
Available Assistance:
We can provide help if you have any questions regarding this Correction.
For any health care professional with medical concerns, please contact Cordis to speak to a clinician at 1-800-327-7714 Option 3, Monday through Friday from 9:00 AM to 5:00 PM EST.
For questions related to the Correction that are not adequately addressed in this letter, please contact the Cordis Field Action Coordinator Hal Baden at (786) 313-2365.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution - US Nationwide in the states of MI, AL, AR, CA, CO, FL, GA, IL, IA, LA, MS, NJ, NC, OH, OK, OR, PA, TN, TX, VA, WV, WI, MI, NY, MA, NV, RI, IN, WA, AZ, DC, Puerto Rico and the country of Canada.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.