FS-OMNI, Fusion Omni-Tome, Sterile, Rx only, Wilson Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105, EC Representative: Cook Ireland Ltd, O'Haioran Road, National Technology Park, Limerick, Ireland. The Fusion Omni-Tome is supplied sterile and is a single use device. This device is used in cannulation of the ductal system and for sphincterotomy.

FDA Recall Z-0325-2009 · Posted 2009-01-05

Recalling FirmCook Endoscopy
Recall StatusTerminated
Product CodeKNS
Date Initiated2008-10-30
Date Terminated2009-04-07
Product Quantity11

Reason for Recall

The wire control port may have been omitted during manufacture.

Firm Action

Cook Ireland (distributor) was contacted via email on 10/30/08 and was instructed to contact their customers and coordinate the return of all recall product to Cook Endoscopy. A response sheet was attached. Recall status reports will be issued monthly. All devices returned as part of this recall initiative will be dispositioned to scrap and destroyed. Should you have any questions, please contact Cook Endoscopy Customer Quality Assurance Department at 1-336-744-0157.

Root Cause

Process control

Distribution Pattern

Ireland