Terminated

Mac Pin Non-Cannulated Screw, 6.5 x 60mm

FDA Recall Z-0325-2018 · Posted None

Recalling FirmAmendia, Inc
Recall StatusTerminated
Product CodeNKB
Date Initiated2017-07-07
Date Terminated2019-10-15
Product Quantity41 screws

Reason for Recall

Incorrect labeling for the lot of MAC Pins. Although the pins are laser as 460 mm in length, the pins are 60 mm in length.

Firm Action

Customers were sent recall notification letters on 07/07/2017. Instructions included to examine inventory and quarantine affected products, notify customers if affected products were further distributed, coordinate the return of affected products back to Amendia, and to complete and return the questionnaire. If there are questions, Amendia Customer Service can be contacted at 770-575-5224.

Root Cause

Process control

Distribution Pattern

Distribution to Georgia, USA.