Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.
FDA Recall Z-0326-2017 · Posted 2016-10-31
Advertisement
Recalling FirmHill-Rom, Inc.
Recall StatusTerminated
Product CodeFSA
Date Initiated2016-08-09
Date Terminated2017-03-24
Product Quantity188
Reason for Recall
Ceiling Bracket 71 was
shipped with a larger center hole. A potentially
hazardous situation exists that if the Ceiling Bracket 71
with a larger center hole is used for a concrete ceiling
Ultra System installation with the M8 screw and the safe
working load is exceeded, the system could fall.
Firm Action
On 8/9/2016, URGENT: MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier.
The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. If you have any questions regarding this Field Safety Correction, please contact Hill-Rom Technical Support at 1-800-445-3720, your distributor or your Hill-Rom representative.
Root Cause
Mixed-up of materials/components
Distribution Pattern
WA NC OH PA MI Australia Sweden
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.