Roche brand OMNI 8 clinical chemistry analyzer; catalog numbers GD0455 and GD0455R.
FDA Recall Z-0327-04 · Posted 2004-01-10
Advertisement
Recalling FirmRoche Diagnostics Corp.
Recall StatusTerminated
Product CodeJJE
Date Initiated2003-12-15
Date Terminated2004-08-02
Reason for Recall
If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume.
Firm Action
Consignees were notified by letter mailed on or about December 15, 2003. Customers are instructed to assure an adequate sample volume until new software becomes available.
Root Cause
Other
Distribution Pattern
United States.