Axiom Luminos dRF without ST filter. . Solid state x-ray imager (flat panel/digital imager). Model number 10252048. Universal fluoroscopic x-ray diagnostic system.

FDA Recall Z-0327-2009 · Posted 2008-11-28

Recall StatusTerminated
Product CodeMQB
Date Initiated2008-09-19
Date Terminated2009-09-23
Product Quantity1 unit

Reason for Recall

Issue may occur where an intended movement of the joystick may unintentionally initiate an x-ray exposure.

Firm Action

The recalling firm issued Update Instruction AX042/08/S dated August 2008, to affected customers. This Update instruction provided Siemens service engineers with instructions for the replacement of the joystick module at each facility.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution --- including states of KY, NJ and PA.