VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
FDA Recall Z-0327-2026 · Posted 2025-10-24
Advertisement
Recalling FirmCanon Medical System, USA, INC.
Recall StatusOpen, Classified
Product CodeLNH
Date Initiated2025-10-16
Product Quantity2 units
Reason for Recall
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
Firm Action
Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form
Root Cause
Device Design
Distribution Pattern
US distribution to Ohio and Nevada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.