Roche brand OMNI 9 clinical chemistry analyzer; catalog numbers GD0475, GD0475R and GD047591.

FDA Recall Z-0328-04 · Posted 2004-01-10

Recall StatusTerminated
Product CodeJJE
Date Initiated2003-12-15
Date Terminated2004-08-02

Reason for Recall

If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume.

Firm Action

Consignees were notified by letter mailed on or about December 15, 2003. Customers are instructed to assure an adequate sample volume until new software becomes available.

Root Cause

Other

Distribution Pattern

United States.