32CM ALTA EXCHANGE Model: AL27SH32 XP ALTA EXCHANGE KIT 15.5 Fr polyurethane twin lumen fixed tip hemodialysis catheter, cuff and side holes for chronic hemodialysis and apheresis. Latex free

FDA Recall Z-0329-2010 · Posted 2009-11-17

Recall StatusTerminated
Product CodeMSD
Date Initiated2009-10-23
Date Terminated2010-06-09
Product Quantity24 units

Reason for Recall

There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira

Firm Action

Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.

Root Cause

Process control

Distribution Pattern

Worldwide distribution - Nationwide, Government. Foreign: Canada, France.