Terminated

Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and other organic matter during endoscopic procedures.

FDA Recall Z-0329-2014 · Posted 2013-11-15

Recalling FirmOlympus America Inc.
Recall StatusTerminated
Product CodeFEQ
Date Initiated2013-05-20
Date Terminated2016-10-24
Product Quantity263

Reason for Recall

There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mode.

Firm Action

Notification letters were sent to customers on May 20, 2013.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution: USA (nationwide) and the country of Canada.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →