Allura Xper FD20/10, Release 1.2, Philips Medical Systems, Bothell, WA 98021
FDA Recall Z-0330-2008 · Posted 2007-12-07
Advertisement
Recalling FirmPhilips Medical Systems North America Co. Phillips
Recall StatusTerminated
Product CodeIZI
Date Initiated2007-09-24
Date Terminated2012-05-03
Product Quantity53 units
Reason for Recall
Inaccurate anatomical measurements when using the optional SmartMask feature
Firm Action
The recalling firm sent customers an Urgent Device Correction notification letter dated 09/21/07 via certified mail - return receipt requested. This notification provided a short term work around for the device defect. For a permanent fix, the recalling firm is developing software update to check for mismatch at the start of a subtracted fluoroscopy run, and will contact customers to schedule an install. Installation is expected to begin 10/2007 and be completed 06/2008.
Root Cause
Device Design
Distribution Pattern
Nationwide