External Compression Device; 1 unit per package Stryker Trauma GmbH; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 Active Apposition/compression is achieved by threading the External Compression Device into the Nail holding Screw. While being threaded into the Nail Holding Screw the tip of the External Compression Device will contact the (shaft screw(cross screw) and will push the screw in distal direction. As the cross screw is screwed into the bone of the patient, the bone fragment will be moved too is distal direction. Finally both sides of the bone fracture will be pressed together - compression is now applied to the fracture side.

FDA Recall Z-0330-2011 · Posted 2010-11-09

Recall StatusTerminated
Product CodeHSB
Date Initiated2010-08-30
Date Terminated2012-08-21
Product Quantity11

Reason for Recall

Variations in the outer diameter of the shaft of the External Compression Device and the inner diameter of the Nail Holding Screw may make assembly of these 2 instruments not possible.

Firm Action

Stryker sent Recall notification letters by Federal Express on September 9, 2010 to all Stryker branches/agencies and Risk Managers, Chief of Orthopaedics of affected hospitals and Surgeons on September 10, 2010, that used the product. The letter identified the product, the problem, and the action that customers should take. Customers were instructed to: Examine their inventory and hospital locations to identify the affected product and return it to their branch or agency warehouse for reconciliation. Reconcile all product on the attached Product Recall Acknowledgment Form and fax a copy to (201) 831-6069 within 5 days of receipt of notice. A Stryker representative would contact the customer with details for return of the product. For any questions regarding this recall call (201) 972-2100.

Root Cause

Other

Distribution Pattern

Nationwide Distribution including AZ, CA, CO, FL, GA, IL, KY, MD, OH, OK, and TX