Brilliance iCT Computed Tomography X-Ray System, Product Usage: The Brilliance iCT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

FDA Recall Z-0335-2014 · Posted 2013-11-15

Recall StatusTerminated
Product CodeJAK
Date Initiated2013-10-11
Date Terminated2015-08-04
Product Quantity9 units

Reason for Recall

If the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with the user due to the length of the accessory.

Firm Action

Philips sent an Urgent - Medical Device Correction letter dated October. 11, 2013, to their customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue the use of the footrest extension. Philips also, stated they will replace the old footrest extension with the current one in use. For questions contact your local Philips representative or local Philips Healthcare office.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution: USA (nationwide) Distribution in the states of DC, GA, IL, IN, NY and OH and the countries: Germany, Israel and Netherlands.