Terminated

CAIRE Liberator 45, MODEL 13262253, Liquid Oxygen System Unit

FDA Recall Z-0337-2023 · Posted 2022-12-02

Recalling FirmCaire, Inc.
Recall StatusTerminated
Product CodeBYJ
Date Initiated2022-10-10
Date Terminated2024-08-19
Product Quantity70 devices

Reason for Recall

An audit discovered some inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle of the device on some of the units.

Firm Action

The firm issued an URGENT: MEDICAL DEVICE RECALL notice its consignees be email and telephone. The notice explained the issue and requested the removal of the affected products from the field and return to CAIRE. CAIRE will immediately replace the above serial numbers with new Liberator devices. Please contact CAIRE Customer Service via phone at 1-800-482-2473 or 770-721-7700 or via email at [email protected] to arrange for the return and replacement of the above serial numbers.

Root Cause

Process control

Distribution Pattern

US Nationwide - Worldwide Distribution: CA, NC, OH, AZ, and Canada, Chile, Colombia, and Germany