Hill-Rom brand Century+ bed; model P1400.

FDA Recall Z-0338-04 · Posted 2004-01-10

Recalling FirmHill-Rom, Inc.
Recall StatusTerminated
Product CodeFNL
Date Initiated2003-12-26
Date Terminated2004-05-13
Product Quantity222

Reason for Recall

If fluid is spilled on the siderail or the bed control areas, the bed may change positions without user input.

Firm Action

A recall letter dated December 26, 2003 was sent to each customer. Customers are asked to follow instructions in the user manual, if there is a fluid spill onto a unit, until Hill-Rom can inspect each unit to determine whether it is affected by the problem.

Root Cause

Other

Distribution Pattern

United States, Canada , Panama and Taiwan.