Hill-Rom 100 Low Bed, Model P3930. AC-powered adjustable hospital bed.
FDA Recall Z-0338-2010 · Posted 2009-11-18
Advertisement
Recalling FirmHill-Rom, Inc.
Recall StatusTerminated
Product CodeFNL
Date Initiated2009-04-20
Date Terminated2010-10-15
Product Quantity4017 beds
Reason for Recall
The siderails may exhibit false latching, or unintentional lowering, or may become inoperable.
Firm Action
Hill Rom notified consignees by recall letter dated 4/20/09 and advised them of the problem and requested them to locate the beds. Consignees with patients in need of siderails were directed to be moved to a different model bed with siderails and, if the patient did not need siderails; consignees were requested to remove the siderails, move the siderails to a stowed position, to tie the siderail to the bed frame or to tag the siderail as out of service. The letter stated the consignee would be contacted within 90 days by a firm representative to do a correction on all affected units.
Root Cause
Process control
Distribution Pattern
Worldwide distribution: USA, Canada, France, India, Singapore and Spain.