NeuroSensor Cerebral Blood Flow and Intracranial Pressure Monitoring Kit
FDA Recall Z-0339-06 · Posted 2006-01-06
Advertisement
Recalling FirmIntegra Neuro Sciences
Recall StatusTerminated
Product CodeGWM
Date Initiated2005-09-08
Date Terminated2012-03-07
Product Quantity79
Reason for Recall
Certain Model NS-P probes contained in the NeuroSensor Cerebral Blood Flow and Intracranial Pressure Monitoring Kit may allow an Intracranial Pressure (ICP) drift of up to 5 mmHg during the first 24 hours of monitoring ICP. This is outside the products labeling specification of 2 mmHg maximum ICP drift within the first 24 hours of monitoring.
Firm Action
Notify Customers and Sales Representatives who received the product by letter sent via Federal Express. Customers will be instructed that an Integra Sales Representative will contact them to arrange for return of affected product.
Root Cause
Other
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.