Terminated

Bivona Adult Tracheostomy Tube, I.D. 8.5 mm, O.D 11.0 mm, Length 88 mm, TTS 11.8 mm, Mfrd by Smiths Medical Critical Care; Product code 670185.

FDA Recall Z-0339-2007 · Posted 2007-01-05

Recall StatusTerminated
Product CodeJOH
Date Initiated2006-08-31
Date Terminated2007-05-04
Product Quantity197

Reason for Recall

Mislabeled- The outer diameter (OD) of the shaft is labeled as 11.0 m on the inner, outer and tray label in error. The OD is actually 11.8 mm.

Firm Action

Distributors and hospital consignees were notified via recall letter dated 8/25/06 to return all unused product to the firm. Distributors were asked to notify their customers of the recall.

Root Cause

Other

Distribution Pattern

Worldwide Distribution ---- including USA states of Arkansas, California, Georgia, Illinois, Kansas, Minnesota, Missouri, Ohio, Pennsylvania, Texas and countries of Canada and Netherlands.