Philips Heartstart MRx Monitor/Defibrillator Models: M3535A (Hospital Use) , M3536A (EMS Use)

FDA Recall Z-0341-05 · Posted 2004-12-21

Recall StatusTerminated
Product CodeMKJ
Date Initiated2004-12-15
Date Terminated2016-02-02
Product Quantity2520 units

Reason for Recall

In Software versions below A.02.00, with the sync function enabled and the device switched from manual to AED, there may be an unexpected delay in delivery of energy

Firm Action

Philips notified customers on 12/13/04 by letter. Users are provided with action required to Mitigate Risk' until the correction is made by a Systems representative.

Root Cause

Other

Distribution Pattern

Nationwide Foreign:Canada, Austria, Australlia, Belgium, China, Denmark, Finland, France, Germany, Israel, Italy,Kuwait, Latvia, Lebannon, Libya, Netherland,Norway, Portugal, Qatar, Russia, Singapore,South Africa, Spain, Switzerland, Sweden, Taiwan, Thailand,Tunisia, UK, United Arab Emirates, Yemen