Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Silicone Central Venous Catheter with PASV Valve and Polyester Cuff - 6.6F. Boston Scientific Cat. #CVC661IK M001CVC661IK0. Firm on label: Boston Scientific Corporation, NAMIC Technology Center, Glens Falls, NY 12801.

FDA Recall Z-0341-06 · Posted 2006-01-06

Recall StatusTerminated
Product CodeLJS
Date Initiated2005-10-28
Date Terminated2007-02-15
Product Quantity434 units

Reason for Recall

Device kits contain B. Braun Locking Tearaway Introducer Systems labeled as 7Fr but may contain 8 Fr sheaths.

Firm Action

Boston Scientific letter dated 10/28/2005 sent via Federal Express. Customers (hospitals) instructed to remove and return product.

Root Cause

Other

Distribution Pattern

DOMESTIC: Nationwide distribution. FOREIGN: Spain, France, Italy, Great Britain, Sweden, Jordan

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑