TC pCO2 Electrodes, Model/Part #945-655-E5260, 510(k) #K043003. Product Usage: The device is intended for used with the firm's TCM series monitor. The E260mpCO2 electrode is a non-invasive device, used with the TCM series monitor to measure transcutaneous pCO2 (through the skin).

FDA Recall Z-0341-2012 · Posted 2011-12-01

Recall StatusTerminated
Product CodeLKD
Date Initiated2011-09-27
Date Terminated2012-02-19
Product Quantity4 units were distributed.

Reason for Recall

Transcutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all calibrations and never get ready for use.

Firm Action

Radiometer contacted the affected U.S. customer and informed them of the recall via telephone on September 27, 2011. Radiometer requested the return of the affected product received from the firm. For questions call (440) 871-0463.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution - USA (nationwide) in the state of San Diego, CA.