Philips HeartStart XL, Model M4735A Defibrillator/Monitor

FDA Recall Z-0343-03 · Posted 2002-12-18

Recall StatusTerminated
Product CodeMKJ
Date Initiated2002-11-25
Date Terminated2012-05-03
Product Quantity260 units

Reason for Recall

Defective energy switch may cause Defibrillator to turn off or charge to a different energy level than was selected.

Firm Action

On 11/25/02 Philips notified customers by letter and requested users to identify the device and use an alternate defibrillator. Philips systems representatives will repair the device in the field. Instructions, titled: 'Risk Mitigation Procedure,' were provided if the unit cannot be pulled from service.

Root Cause

Other

Distribution Pattern

AL, AZ, CA, CT, DE, FL, IL , IN, LA, NE, NY, PA, TX Nationwide Foreign: Canada, Algeria, Belgium, France, Great Britain, Ireland, Israel, Italy, Lebanon, Spain, Sweden, Hong Kong, Philippines, Singapore, Taiwan, China, Japan,Trinadad