SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured & Distributed 12/29/2006. Version 4.0.1.15 Manufactured & Distributed 05/02/2007. Version 4.0.1.16 Manufactured & Distributed 08/22/2008. Used in a medical research or clinical laboratory setting by knowledgeable, trained and experienced personnel.

FDA Recall Z-0344-2011 · Posted 2010-11-10

Recalling FirmSCC Soft Computer
Recall StatusTerminated
Product CodeJQP
Date Initiated2010-02-16
Date Terminated2012-07-17
Product Quantity31

Reason for Recall

When a test is being verified, the wrong (expired) ranges may be associated to that test result, causing an incorrect flag to be posted for that test result.

Firm Action

Clients were notified via the proprietary communication software on 02/16/2010. The Notification Verbiage identified the affected product and informed clients of the potential hazard, root cause, corrective action, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.

Root Cause

Software design

Distribution Pattern

Nationwide Distribution - US and Puerto Rico.