Terminated

AMS600-B, AMS700-B, SA-0105, SA-3675, SA-3678 Arteriocyte Medical systems, Magellan, Cannula Tip, Catalog Number AMS600-B,Sterilized by Ethylene Oxide, Rx Only, Manufacturer Micromedics Inc., St. Paul, MN 55121-1385 USA, Distributor: Arteriocyte Medical Systems, Inc., 45 South Street, Hopkinton, MA 01748 USA. Arteriocyte Medical systems, Magellan, Cannula Tip, Catalog Number AMS700-B,Sterilized by Ethylene Oxide, Rx Only, Manufacturer Micromedics Inc., St. Paul, MN 55121-1385 USA, Distributor: Arteriocyte Medical Systems, Inc., 45 South Street, Hopkinton, MA 01748 USA. Fibrijet, Applicator Tip Dual Cannula Malleable 26 ga x3" (7.6cm), Catalog Number SA-0105, Micromedics, Sterilized by Ethylene Oxide, Rx Only, Micromedics Inc. St Paul, MN 55121-1385 USA. Fibrijet, Mixing Applicator Tip Low Viscosity with Spray Tip, Catalog Number SA-3675, Micromedics, Sterilized by Ethylene Oxide, Rx Only, Micromedics Inc. St Paul, MN 55121-1385 USA. Fibrijet, Blending Connector with Mixer, Catalog Number SA-3678, Micromedics, Sterilized by Ethylene Oxide, Rx Only, Micromedics Inc. St Paul, MN 55121-1385 USA

FDA Recall Z-0345-2011 · Posted 2010-11-18

Recalling FirmMicromedics, Inc.
Recall StatusTerminated
Product CodeKYZ
Date Initiated2010-03-15
Date Terminated2011-06-21
Product Quantity4025 (4295 pouches)

Reason for Recall

Micromedics, Inc is conducting a product recall on a number of medical device products because of weak seals of the sterile pouches, potentially resulting in a non-sterile product which may cause transmission of disease or infection.

Firm Action

Micromedics issued a Product Recall letter dated March 15, 2010 to customers, describing the product being recalled, the product sterility issue, and actions to be taken. The letter advised customers to contact Customer service to return any affected product to Micromedics for replacement. Customers can contact Micromedics at 1-800-624-5662 or 651-452-1977 ext. 226 concerning this recall.

Root Cause

Packaging process control

Distribution Pattern

Nationwide Distribution in the states of CA, CT, DE, FL, GA, ID, IL, IN, MD, MA, MI, MN, NJ, NY, NC, PA, TX, UT, and WI. FRANCE, NETHERLANDS, BELGIUM, UNITED KINGDOM, CANADA, ISRAEL, PERU, SPAIN, IRELAND.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →