Terminated

UniCel Dxl 800 Access Immunoassay System, Part Numbers: 973100, A71456, and A71457. The UniCel DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel DxC series of instruments is a DxI Immunoassay System integrated with a chemistry analyzer.

FDA Recall Z-0345-2013 · Posted 2012-11-19

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeJJE
Date Initiated2011-08-01
Date Terminated2012-11-19
Product Quantity1,756 units

Reason for Recall

The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.

Firm Action

Beckman Coulter sent a field modification was released in August 2011 which included: - Installing a new aspirate and dispense tubing routing bracket. - Replacing existing aspirate and dispense probe tubing with modified tubing to help eliminatemisrouting of aspirate and dispense probe tubing. During installation the FSE informed the customer: -That aspirate and dispense tubing are different lengths. - That probes are not interchangeable with other instruments once tubing has been modified. - To route probe tubing by referring to the labels on the tubing. An updated service document, Routine Maintenance Changes, A82294B was provided to every customer. For questions regarding this recall call 714-961-4483.

Root Cause

Process design

Distribution Pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, AE, AU, BG, BW, CH, CN, CZ, DE, ES, FR, GB, HK, HR, HU, IN, IT, JO, JP, LB, MA, MX, NL, PL, QA, RO, RU, SK, TN, TR, ZA, AO, AR, AT, BE, CO, DZ, FI, IE, IL, KR, KW, LI, NA, NO, NZ, PF, PH, PR, SE, SG, SZ, TH, TW, VE, VN, and YT.