Hill-Rom CareAssist Bed; model P1170.
FDA Recall Z-0346-05 · Posted 2004-12-22
Advertisement
Recalling FirmHill-Rom, Inc.
Recall StatusTerminated
Product CodeFNL
Date Initiated2004-11-29
Date Terminated2005-06-02
Product Quantity1,825
Reason for Recall
Warning label containing instructions for proper electrical grounding was not attached.
Firm Action
Consignees were notified by letter dated 11/29/04 and provided labels to affix to their beds.
Root Cause
Other
Distribution Pattern
Nationwide, Canada and Saudi Arabia.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.