7F Locking Tearaway Introducers non-sterile. Part number 614006. The product is packaged as 50 units per pouch, 2 pouches per carton.
FDA Recall Z-0346-06 · Posted 2006-01-06
Advertisement
Recalling FirmB. Braun Medical, Inc.
Recall StatusTerminated
Product CodeDYB
Date Initiated2005-09-01
Date Terminated2006-01-05
Product Quantity8000 units
Reason for Recall
Tear away sheath assembly may contain a sheath (8FR) that is not the correct French size for the dilator (7FR).
Firm Action
The recalling firm telephoned their customers on 91/05 & 9/14/05 to inform them of the problem.
Root Cause
Other
Distribution Pattern
The product was shipped to two consignees that repack the product into their own kits.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.