Axiom Artis System, X-ray, Angiographic

FDA Recall Z-0347-03 · Posted 2002-12-18

Recall StatusTerminated
Product CodeIZI
Date Initiated2002-08-05
Date Terminated2004-10-25
Product Quantity134

Reason for Recall

Set delay time for contrast injection exceeds 10 seconds, radiation release button is blocked.

Firm Action

A Safety Advisory Letter was sent 7/23/2002 by Siemens Medical. The letter advised the technicians of the problem and Siemens would be coming out to do the software modificaitons.

Root Cause

Other

Distribution Pattern

The device was distributed to Hospitals nationwide.