Omnipod DASH Insulin Management System, Product Catalog Number 18239 M/D: PDM USA1 D001 MG USA1
FDA Recall Z-0347-2020 · Posted None
Advertisement
Recalling FirmInsulet Corporation
Recall StatusTerminated
Product CodeLZG
Date Initiated2018-12-18
Date Terminated2020-05-13
Product Quantity264
Reason for Recall
There is a potential for a communication interruption following a bolus command that may result in inaccurate information presented in insulin on board (IOB), last bolus field or bolus history.
Firm Action
Urgent Medical Device Correction letters dated 12/18/18 were sent to customers.
Root Cause
Software design
Distribution Pattern
US Nationwide.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.