Terminated

Platelia Aspergillus EIA

FDA Recall Z-0348-04 · Posted 2004-01-14

Recall StatusTerminated
Product CodeNOM
Date Initiated2003-11-06
Date Terminated2004-01-27
Product Quantity147 kits

Reason for Recall

Potential drug and device interaction with piperacillin/tazobactam (Zosyn) causing positive test result for galactomannan.

Firm Action

Customers were sent a letter, Technical Bulletin and revised package insert on November 6, 2003.

Root Cause

Other

Distribution Pattern

The firm distributed to hospitals and diagnostic laboratories nationwide.