Arthrocare 3mm ParaFix Anchor, single strand of Magnum Wire,Part Number: 22-5010, Axya Medical, Inc., Beverly, MA 01915
FDA Recall Z-0348-2008 · Posted 2007-12-08
Advertisement
Recalling FirmAxya Medical
Recall StatusTerminated
Product CodeMBI
Date Initiated2007-09-27
Date Terminated2012-06-21
Product Quantity338 units
Reason for Recall
Sterility Compromised (Package Integrity): Sterility of the device may be compromised due to holes in the sterile packaging
Firm Action
Axya Medical notified customers by letter, Urgent-Product Recall, dated 9/25/07 , via UPS . Users are requested to return product and complete the reply form.
Root Cause
Package design/selection
Distribution Pattern
Nationwide