Terminated

Candela Laser GentleLase Pro, Powered Laser Surgical Instrument for dermatological use - Model number: 9914-00-9015. and Candela Laser GentleLase Pro LE, Model number: 9914-00-9040

FDA Recall Z-0348-2014 · Posted 2013-11-18

Recalling FirmCandela Corporation
Recall StatusTerminated
Product CodeGEX
Date Initiated2013-11-06
Date Terminated2015-06-08
Product Quantity1258 units

Reason for Recall

Unintended single pulse maybe emitted to handpiece prior to pressing of trigger or footswitch and cause injury.

Firm Action

Candela notiifed Distributors on 11/8/13 through e-mail, Certified return receipt request, or FedEx package containing distributor letter, customer letter for translation and customer lists. A software update to Version 2.0 will be issued to the user by a field service representative.

Root Cause

Software design

Distribution Pattern

Worldwide distribution: US (nationwide) and Internationally to: Azerbaijan, Bulgaria, Chile, Czech Republic, Greece, Israel, Kuwait, Qatar, Saudi Arabia, Serbia, Turkey, and the United Arab Emirates and to Subsidiaries in Australia, France, Germany, Italy, Japan, Portugal, Spain, and the United Kingdom.