ECG Out Cable - Heartstart MRx M1783A/M5526A 12-pin Sync cables All units of Philips sync cables model M1783A and M5526A as well as Philips ECG out cables model 989803195641. Manufactured - September 4, 2002 October 8, 2014
FDA Recall Z-0348-2018 · Posted None
Advertisement
Recalling FirmPhilips Electronics North America Corporation
Recall StatusTerminated
Product CodeMKJ
Date Initiated2017-07-17
Date Terminated2018-07-12
Product Quantity15811
Reason for Recall
Issue with ECG out cables. When a Philips monitor/defibrillator is receiving an ECG signal from an auxiliary bedside monitor via a sync cable or ECG out cable, the following can occur if the monitor/defibrillator experiences interference from electrical fast transients (EFTs):
" On the HeartStart MRx and HeartStart XL, EFT noise may be misinterpreted as an R-wave.
" On the HeartStart XL+, EFT noise can disable ECG monitoring and potentially interrupt demand mode pacing*.
*Note: It is contrary to the XL+ Instructions for Use to perform demand mode pacing while using the ECG out cable or obtaining the ECG signal from a bedside monitor. The XL+ Instructions for Use includes the following warning: When pacing in Demand Mode, the ECG cable from the patient must be directly connected to the HeartStart XL+. If the user follows this warning, this problem cannot occur on the XL+.
Firm Action
Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent
Medical Device Correction notification/Field Safety Notice:
Customers should remove affected sync cables and ECG out cables from service and destroy them, documenting such action on the attached reply form, which should be returned to Philips. Customers should retain a copy of the reply form.
The customer should follow the Instructions for Use that describe the recommended approach for performing cardioversion with the XL+ and MRx monitor/defibrillators. The ECG trunk cable can be disconnected from the Philips bedside monitor and connected to the monitor/defibrillator before delivering therapy; this is the preferred source of an ECG waveform for synchronization.
Root Cause
Device Design
Distribution Pattern
1067 US customers 163 Canada, 917 International
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K062233HEARTSTART MRX MONITOR/DEFIBRILLATORPhilips Medical Systems North America, Inc. · 2006-11-22K061707HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONSPhilips Medical Systems · 2006-08-11K051134HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536APhilips Medical Systems · 2005-09-08K040404HEARTSTART MRX, MODEL M3535APhilips Medical Systems, Inc. · 2004-05-04K031187HEARTSTART MRX, MODEL M3535APhilips Medical Systems · 2003-08-27
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.