Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

GREER DIA - KIT ANTIGEN M. FAENI, 2.0mL 20 MG/ML, Analyte Specific Reagent, Item: K62, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.

FDA Recall Z-0348-2023 · Posted 2022-11-30

Recall StatusTerminated
Product CodeDGW
Date Initiated2022-10-25
Date Terminated2026-02-11
Product Quantity102 units

Reason for Recall

The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.

Firm Action

The firm initiated the recall beginning on 10/25/2022 by telephone and email, following with a "MEDICAL DEVICE RECALL" notice. The notice stated the problem and requested the following: "Greer requests that you immediately discontinue use of the M. faeni Dia-Kit antigen lot(s) and return any remaining stock of the lots in question. Please contact a Greer Customer Service Representative at 1-800-378-3906 between 8 a.m. and 5 p.m. Eastern Time to receive a Returned Goods Authorization number and arrange for return of the product. Your Greer Customer Service Representative will also arrange for replacement of the product."

Root Cause

Employee error

Distribution Pattern

US Distribution to states of: CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑